🔴 Sound Level Meter — Calibration & Service Log
← Back to list
⚠ Use a Class 1 or Class 2 acoustic calibrator. Calibrate in a quiet area away from HVAC noise. Take a photo of the SLM with the calibrator attached before starting.
Job Details
Mobile / Unit *
GM1
GM2
GM3
GM4
GM5
GM6
GM7
GM8
SIOC 2
SIOC 3
SIOC 4
RAF
Guud Vision Clinic
Cape Town
Service Date *
Location
Technician First Name *
Technician Surname *
Machine Brand
Machine Model
Serial Number
Acoustic Calibrator Brand/Model
Calibrator Serial Number
Before You Start — Required Photo
Photo — SLM with Calibrator (Before) *
Damage Assessment
Component
Condition
Comments
Microphone capsule — check for dents, contamination
OK
Minor Wear
Damaged
N/A
Windscreen / foam cover condition
OK
Minor Wear
Damaged
N/A
External housing — cracks, loose buttons
OK
Minor Wear
Damaged
N/A
Display screen
OK
Minor Wear
Damaged
N/A
Battery compartment and contacts
OK
Minor Wear
Damaged
N/A
Measurement cable (if external microphone)
OK
Minor Wear
Damaged
N/A
Storage case condition
OK
Minor Wear
Damaged
N/A
Service Checklist
Done
Task
Comments
Clean microphone capsule surface (no pressure, dry lint-free cloth)
Inspect and replace windscreen if torn or saturated
Clean housing and screen
Check battery level and contacts for corrosion
Test A-weighting filter response
Test C-weighting filter response
Verify Fast and Slow time weightings
Check memory/data-logging function (if applicable)
Update firmware (if applicable)
Calibration Record
Step
Method / Target
Result
Comments / Measured Value
Acoustic Calibrator Check (94 dB @ 1 kHz)
Insert calibrator; adjust ref level to 94.0 dB; record actual reading
Pass
Fail
A-weighted Level (dBA)
Measure calibrator output with A-weighting; target 94 dB ± 0.5 dB
Pass
Fail
C-weighted Level (dBC)
Measure calibrator output with C-weighting; target 94 dB ± 0.5 dB
Pass
Fail
Fast Time-Weighting
Confirm Fast integration responds to impulsive signal
Pass
Fail
Slow Time-Weighting
Confirm Slow integration averages correctly
Pass
Fail
Reference Level Verification
Compare reading to accredited calibrator; record deviation
Pass
Fail
Post-calibration acoustic check
Re-insert calibrator; final reading ± 0.5 dB from 94 dB
Pass
Fail
Post-Service Functionality Check
Result
Check
Comments
Pass
Fail
Readings within ±0.5 dB of acoustic calibrator
Pass
Fail
A and C weighting filters functional
Pass
Fail
Fast and Slow time weightings responding correctly
Pass
Fail
Windscreen clean and in good condition
Pass
Fail
Battery fully charged / replaced
Pass
Fail
Calibration certificate attached (if sent to accredited lab)
Pass
Fail
Next calibration due date recorded
After Service — Required Photo
Take the same photo as before to confirm the SLM and microphone are clean and reassembled.
Photo — SLM (After) *
Notes & Sign-off
Technician Notes
Technician Signature (type full name)
Technician Sign-off Date
Supervisor Signature (type full name)
Supervisor Sign-off Date
Post-Service Photos / Calibration Certificate (optional)
Save Calibration Log