🎤 Audiometer — Calibration & Service Log
← Back to list
⚠ Calibrate in a quiet room (ambient noise < 40 dB SPL). Take a clear photo of the audiometer before starting. Use a calibrated SPL meter or acoustic calibrator where specified.
Job Details
Mobile / Unit *
GM1
GM2
GM3
GM4
GM5
GM6
GM7
GM8
SIOC 2
SIOC 3
SIOC 4
RAF
Guud Vision Clinic
Cape Town
Service Date *
Location
Technician First Name *
Technician Surname *
Machine Brand
Machine Model
Serial Number
Before You Start — Required Photo
Photo — Audiometer (Before) *
Damage Assessment
Component
Condition
Comments
External housing — check for cracks, dents, or damage
OK
Minor Wear
Damaged
N/A
Headphones / supra-aural earphones — cups, cushions, headband
OK
Minor Wear
Damaged
N/A
Insert earphones — ear tips, tubing, connectors
OK
Minor Wear
Damaged
N/A
Bone vibrator — transducer, cushion, headband
OK
Minor Wear
Damaged
N/A
All control knobs, dials, and buttons
OK
Minor Wear
Damaged
N/A
Signal cables and connectors (fraying, bent pins)
OK
Minor Wear
Damaged
N/A
Power supply / battery compartment
OK
Minor Wear
Damaged
N/A
Display screen or reading scale
OK
Minor Wear
Damaged
N/A
Service Checklist
Done
Task
Comments
Clean external housing, knobs, and dials
Inspect and clean headphone cushions; replace if worn
Check insert earphone tips; replace if damaged or discoloured
Inspect bone vibrator cushion and headband
Test output at all frequencies: 125, 250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz
Verify masking noise (narrowband and speech noise) output
Check all attenuation steps (5 dB steps across full range)
Verify channel 1 and channel 2 (if applicable)
Update firmware or test-tone calibration file (if applicable)
Document ambient noise level at time of calibration
Calibration Record
Step
Method / Target
Result
Comments / Measured Value
1 kHz Reference Level
Set to 60 dB HL — verify output with SPL meter; target ±1 dB
Pass
Fail
Frequency Accuracy — 250 Hz
Verify tone frequency with frequency counter; target ±1%
Pass
Fail
Frequency Accuracy — 500 Hz
Verify tone frequency; target ±1%
Pass
Fail
Frequency Accuracy — 1 kHz
Verify tone frequency; target ±1%
Pass
Fail
Frequency Accuracy — 2 kHz
Verify tone frequency; target ±1%
Pass
Fail
Frequency Accuracy — 4 kHz
Verify tone frequency; target ±1%
Pass
Fail
Frequency Accuracy — 8 kHz
Verify tone frequency; target ±1%
Pass
Fail
Linearity Check
Step attenuator from 20 to 80 dB HL in 10 dB steps; verify linear output
Pass
Fail
Harmonic Distortion (1 kHz)
Total harmonic distortion < 2.5% at 60 dB HL
Pass
Fail
Cross-talk (if dual-channel)
Verify < -70 dB cross-talk between channels
Pass
Fail
Masking Noise Level
Verify narrowband masking at 1 kHz centre frequency ±3 dB
Pass
Fail
Final accuracy check
Compare to certified reference audiometer or calibrated SPL meter
Pass
Fail
Post-Service Functionality Check
Result
Check
Comments
Pass
Fail
All frequencies present with no distortion or dropout
Pass
Fail
Attenuator steps linear within ±1 dB tolerance
Pass
Fail
Headphone / earphone cushions clean and undamaged
Pass
Fail
All controls and buttons fully functional
Pass
Fail
No unusual sounds, clicks, or interference
Pass
Fail
Calibration certificate attached (if sent to accredited lab)
Pass
Fail
Next calibration due date recorded
After Service — Required Photo
Take the same photo as before to confirm the equipment is clean and reassembled.
Photo — Audiometer (After) *
Notes & Sign-off
Technician Notes
Technician Signature (type full name)
Technician Sign-off Date
Supervisor Signature (type full name)
Supervisor Sign-off Date
Post-Service Photos / Calibration Certificate (optional)
Save Calibration Log