🔧 Lensmeter / Vertometer

Service Log #3

← Back to list

Job Details

Mobile / Unit
Guud Vision Clinic
Service Date
2026-07-10
Location
Capt Town
Technician
Nceba Mhawu
Machine Model
COT PLE900
Serial Number
TST10/07/26 VERT

Before You Start — Photos

After Service — Photos

Compare against the "before" photos above to confirm the equipment was cleaned.

Damage Assessment
CheckConditionComments
External Housing OK
Lens Holder / Target Mechanism OK Minor ink spill
Eyepiece Assembly OK
Internal Illumination System OK
Measurement Drum / Control Knob Damaged Lever broken
Power Cable & Plug OK
Display/Reading Scale OK
Service Checklist
TaskDoneComments
Clean external housing and lenses Yes
Inspect and clean eyepiece Yes
Check and lubricate movement of measurement drum Yes
Verify illumination system is functioning Yes
Clean and align lens holder Yes
Calibration of measurement scale Yes
Replace damaged or worn components Yes
Test readings with standard lens for accuracy Yes
Document final calibration values Yes
Calibration Record
StepMethodResultComments
Zero setting of scale Align to 0 with no lens in holder Pass
Sphere calibration Verify with +5.00D and -5.00D test lens Pass
Cylinder calibration Verify with known cylinder lens Pass
Axis calibration Align with lens of known axis Pass
Illumination alignment Adjust light path for clear reticle Pass
Final accuracy test Measure standard lens, compare to spec Pass
Post-Service Functionality Check
CheckResultComments
Optical readings match standard lens tolerance Pass
Illumination brightness adequate Pass
Smooth movement of focusing/measurement controls Pass
No loose parts or rattling Pass
Machine stable on work surface Pass

Notes & Sign-off

Lever needs to be replaced but machine is functioning

Technician Signature
Nceba Mhawu
Technician Sign-off Date
Supervisor Signature
Supervisor Sign-off Date
2026-07-10